Selecting Kilovolts to Reduce the Radiation Dose of CT-Guided Radiofrequency Ablation for Liver Tumors Based on BMI |
ZHAO Junlu, NIE Guanwei, LIU Chong, et al |
The First Hospital, Hebei Medical University, Hebei Shijiazhuang 050031, China |
|
|
Abstract Objective: To explore the effect of selecting kilovolts based on body mass index (BMI) on reducing the radiation dose of radiofrequency ablation in liver tumors under CT guidance. Methods: CT-guided RFA for hepatic ablations performed from January 1, 2019, through December 31, 2019, were retrospectively reviewed (55 men and 45 women; aged, 29~90 years). The radiation dose was recorded during each of the following steps: planning, performing, and postprocedure. BMI-based protocol modification changes in tube voltage and tube current were then applied to hepatic ablations performed subsequently (52 men and 48 women; aged 30~91 years). Image quality and overall operator satisfaction were noted for each case (score, 1~5). Dose reduction after modification was then calculated. Results: Retrospective analysis found a mean (± SD) overall CT dose index (CTDI) for CT-guided RFA to be 12.83±3.78mGy. After protocol modification, the mean CTDI decreased to 3.84 ±2.24 mGy, a 70.07% reduction overall. The image quality was slightly lower compared to before the modification, but the image quality in both stages met the needs of CT-guided radiofrequency ablation for liver tumors. The difference between electrode needle display and operator confidence satisfaction was not statistically significant (P> 0.05). Conclusion: BMI-based kilovolt modifications can significantly reduce the radiation dose during CT-guided percutaneous liver tumor ablation without significantly sacrificing image quality.
|
|
|
|
|
[1] 郭小军,孔延亮,刘涛,等.CT引导经皮射频消融术治疗肝癌分析[J].临床荟萃,2021,36(9):807-810. [2] 王青兵,王雁军,张勇,等.射频消融在老年结直肠癌肝转移治疗中的疗效分析[J].中华肝脏外科手术学电子杂志,2021,10(3):306-309. [3] 司贮元,朱化强,高恒军,等.超声与CT引导下射频消融治疗早期肝细胞癌效果比较[J].中华肝胆外科杂志,2020,26(6):417-421. [4] 刘俊七,程勇,陈澜菁,等.不同CT设备肺结节扫描参数优化的体模对照研究[J].放射学实践,2021,36(12):1577-1581. [5] 赵海波,罗焕江,廖明壮,等.双源CT70与120kV不同CARE Dose 4D预设值下图像质量的对比研究[J].医疗卫生装备,2019,40(2):56-59. [6] 郝俊超,李珍珍.肺部肿瘤低剂量CT引导下射频消融联合化疗临床有效性分析[J].临床医药文献杂志(电子版),2019,6(12):80. [7] 中华医学会放射学分会质量管理与安全管理学组.CT辐射剂量诊断参考水平专家共识[J].中华放射学杂志,2017,51(11):817-822. [8] 房达,崔石昌,李京华,等.CT引导微波消融治疗肝癌中CT剂量的优化控制[J].医学影像学杂志,2016,26(10):1837-1840. [9] 朱功升,肖慧娟,董晓美,等.第3代双源CT低剂量技术在高心率冠脉成像中的应用[J].中国临床医学影像杂志,2021,32(1):52-55. [10] 马桂娜,江奇琦,韩萍,等.SAFIRE重建联合CARE kV技术在主动脉CT低剂量成像中的临床应用[J].临床放射学杂志,2018,37(10):1755-1759. |
|
|
|